正己烷中25种正构烷烃类混标/C9-C33

Reference Material

This certificate is designed in accordance with ISO 17034 and ISO Guide 31. This reference material (RM) was designed,produced and verified in accordance with ISO/IEC 17025ISO 17034 and a registered quality management system ISO 9001.

Product Name 25 N-alkane Mix in n-Hexane,100μg/mL,200μg/mL
Part Number 81447
 
Lot Number 21080493
Expiry Date 2023-Feb-24
 
Storage           (20±5)℃,dark
CERTIFIED
Component
No.
Peak
Sequence
ComponentCAS No.Concentration
(μg/mL)
Uncertainty
(%)
11n-Nonane111-84-21003
22n-Decane124-18-51003
33n-Undecane1120-21-41003
44n-Dodecane112-40-31003
55Tridecane629-50-51003
66n-Tetradecane629-59-41003
77n-Pentadecane629-62-92003
88n-Hexadecane544-76-31003
99n-Heptadecane629-78-71003
1010n-Octadecane593-45-31003
1111n-Nonadecane629-92-51003
1212n-Eicosane112-95-81003
1313n-Heneicosane629-94-71003
1414n-Docosane629-97-01003
1515n-Tricosane638-67-51003
1616n-Tetracosane646-31-11003
1717n-Pentacosane629-99-21003
1818n-Hexacosane630-01-31003
1919n-Heptacosane593-49-71003
2020n-Octacosane630-02-41003
2121n-Nonacosane630-03-51003
2222n-Triacontane638-68-61003
2323n-Hentriacontane630-04-62003
2424n-Dotriacontane544-85-42003
2525n-Tritriacontane630-05-72003

C H R O M A T O G R A M

 
 Instrument
  GC/FID
 Column
  HP-5(30m×0.32mm×0.25μm)
 Injection Temp
  280℃
 Detector Temp
  300℃
 Method Details 

  60℃(1min),8℃/min→290℃,30℃/min→320℃(12min)
Intended Use

This RM is intended for use in a laboratory as a calibration and quality control standard or in method development for analytical techniques.

Uncertainty

The certified value(s) and uncertainty(ies) are determined in accordance with ISO 17034 with an 95% confidence level (k=2). Uncertainty is based on the Total Combined Uncertainty, including uncertainties of preparation, purity of neat materials, homogeneity, stability testing.

Traceability

The balances used for gravimetric measurements are calibrated with weights traceable to the national standards. The calibration of the balances is verified annually by an external accredited calibration service. This analysis method has been verified using an approach consistent with ISO 17034:2016 & ISO 17025:2017.

Homogeneity

Random replicate samples of the final packaged RM have been analysed to prove homogeneity consistent with ISO 17034.

Storage

The RM should be stored in the original sealed bottle at the indicated temperature.